JCM Part No . 960 - 000058R _ Rev . 2 © 2008,   JCM - American Corporation 1 4 Section 4 EBA - 03 Bill Validator Calibration 3 . Click the Screen Button to con - tinue the installation . 4 . To complete the software installation on yo ur hard drive, click the large icon shown in Figure 4 - 6 and follow the screen prompts . Using the Calibration Software With the cabling attached as shown in the Figure 4 - 1 illustration, begin the tests by: 1 . Double - click on the EBA - 03 icon . The open - ing screen shown in Figure 4 - 7 will appear . Click and follow the on - screen instructions . 2 . Once the motor speed test is finished, open th e top of the Acceptor and insert the White Reference Paper (Part No . 074541) into the insertion slot . 3 . Then click on the Screen Button . When that test is finished, replace th e White Reference Paper with the Black Reference Paper (Part No . 074542) . If there is a calibration problem, an ERROR m essage will appear in Figure 4 - 8 indicating where the possible problem location could be . Sensor Names and Locations If an error does occur during the previous tests, use the error information listed in Table 4 - 1 and the location information shown in Figure 4 - 9 to deter - mine which sensor may be at fault and where that particular sens or is located . Figure 4 - 5 Second Software Install Screen Figure 4 - 6 Third Software Install Screen Figure 4 - 7 Initial Adjustment Software Screen Table 4 - 1   Error Codes Error Name Sensor Name l3t _ ir L3DI/L3PT l3t _ gn L3UG/PTL3 l3d _ ir L3DI l3d _ rd L3DR l3u _ ir L3UI l3u _ gn L3UG c0t _ ir C0DI/C0PT c0t _ rd C0UR/PTC0 C0D _ IR C0DI c0d _ rd C0DR c0u _ ir C0UI c0u _ rd C0UR r3t _ ir R3DI/R3PT r3t _ rd R3UR/PTR3 r3d _ ir R3DI r3d _ gn R3DG Figure 4 - 8 Final Adjustment Software Screen NOTE: This test only takes a second . Remove the Black Reference Paper and click “Next” again .